![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 681](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759681.webp)
R&S
®
ZVA/ZVB/ZVT
SENSe
1145.1084.12 6.219
E-6
[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:RPORt <port_no>
Selects the receive port for the harmonic measurement.
<Ch>
Channel number.
<port_no>
Integer port number.
Range [def. unit]
Depending on the analyzer model [–].
*RST value
2
SCPI, Command Types
Device-specific, command or query.
Example:
See [SENSe<Ch>:]FREQuency:CONVersion
[SENSe<Ch>:]FREQuency:CONVersion:HARMonic:SPORt <port_no>
Selects the source port for the harmonic measurement.
<Ch>
Channel number.
<port_no>
Integer port number.
Range [def. unit]
Depending on the analyzer model [–].
*RST value
1
SCPI, Command Types
Device-specific, command or query.
Example:
See [SENSe<Ch>:]FREQuency:CONVersion
SENSe:FREQuency:CONVersion:HARMonic
[SENSe<Ch>:]FREQuency...
This subsystem sets frequency-related parameters, especially the measurement and display ranges for
the different sweep types, and controls the frequency-converting measurement modes (harmonic
distortion, mixer mode, etc.).
The frequency ranges for the different instrument models are listed below:
ZVA8
ZVA24
Start, Stop
300 kHz to 8 GHz
300 kHz to 24 GHz
Center
> 300 kHz to < 8 GHz
> 300 kHz to < 24 GHz
Span
10 mHz to 7,9997 GHz
10 mHz to 23,9997 GHz
Note:
For a frequency sweep the range can be defined alternatively by a combination of Start/Stop
frequencies or a Center frequency and Span. In fixed frequency mode (power sweep or
Time/CW Mode; see [SENSe<Ch/Tr>:]FUNCtion[:ON]) the measurement frequency is set by
means of [SENSe<Ch>:]FREQuency:CW|FIXed.
[SENSe<Ch>:]FREQuency:CENTer <center_frequency>
Defines the center of the measurement and display range for a frequency sweep.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......