R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.211
E-6
External Generators
dialog
before they appear in the list.
•
Take Cal Sweep
initiates a calibration sweep for the selected source. The progress of the
sweep is monitored in the calibration sweep diagram. After the necessary calibrations have
been completed, a message in the trace list indicates the status of the power calibration.
•
Abort Sweep
stops a running sweep. The data acquired in the aborted sweep is discarded.
•
The
Settings
panel displays the essential parameters of the source power calibration, to be
defined in the
Modify Source Power Cal Settings
dialog.
•
Source Power Correction Off
can be selected/deselected as soon as power calibration data
have been acquired for the selected source. If the box is checked, the calibration data is kept in
memory but not applied to the source. The
Power Correction Off
softkey disables the power
correction for all receivers and sources.
•
Modify Settings...
opens the
Modify Source Power Cal Settings
dialog.
•
Power Meter
selects an external power meter that is used to measure the exact source power
at the calibration point. Power meters
(Pmtr1, Pmtr2, ...)
must be configured explicitly in the
System Configuration
–
External Power Meters
dialog
before they appear in the list.
•
Power Meter Config...
opens the
System Configuration
–
External Power Meters
dialog.
•
Perform Verification Sweep after Cal
enables or disables a verification sweep that the analyzer
performs after the source power calibration. The sweep is also displayed in the calibration
sweep diagram. Its purpose is to test the accuracy of the source power calibration after the final
correction data have been acquired. The final power calibration results are applied starting with
the first sweep after the verification sweep.
Remot
e
control
:
SOURce<Ch>:POWer<Pt>:CORRection[:ACQuire]
SOURce<Ch>:POWer<Pt>:CORRection:COLLect[:ACQuire]
SOURce<Ch>:POWer<Pt>:CORRection[:STATe]
SOURce<Ch>:POWer<Pt>:CORRection:GENerator[:STATe]
SOURce<Ch>:POWer<Pt>:CORRection:PMETer:ID
SOURce<Ch>:POWer<Pt>:CORRection:COLLect[:ACQuire]:VERification[:STAT
e]
Calibration Sweep Diagram
The calibration sweep diagram displays the power measured by the external power meter while a
source power calibration is performed.
•
A trace is displayed for each calibration sweep.
•
If
Perform Verification Sweep after Cal
is enabled in the
Source Power Cal
dialog, an additional
trace is displayed after the end of the calibration.
The diagram is automatically scaled so that the center corresponds to the selected
Cal Power.
For
frequency sweeps, a horizontal trace close to the
Cal Power
indicates an accurate power calibration.
For power sweeps, the ideal trace is a power ramp.
Below the sweep diagram, the status messages
Calibration in Progress/Finished
and
Verification in
Progress/Finished
show the progress of the calibration procedure.
Modify Source Power Cal Settings
The
Modify Source Power Cal Settings
dialog displays the target power for the source power calibration
and defines stop criteria.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......