R&S
®
ZVA/ZVB/ZVT
Channel
Menu
1145.1084.12 4.227
E-6
•
Select a standard
Type
and its
Gender
(for polarized/not sexless connector types and if the port
assignment is not restricted) and assign a
Label
.
•
Restrict Port Assignment
•
Select S-Params From
Qualify whether the standard is described by a
Circuit Model
from which the analyzer can
calculate the S-parameters or by a table of measured or simulated S-parameters stored in a
Touchstone file. Pressing the
Read Data from File...
button opens a file selection dialog where
the appropriate file type (*.s1p for one-port standards and *.s2p for two-port standards) is
automatically selected.
The
Sliding Match
and
Attenuation
are special standard types which must be described
by a circuit model. The controls in the
S-Params From
panel are disabled.
For two-port standards described by a *.s2p file, the implicit ports 1 and 2 (given by the order of
S-parameters Re(S11) Im(S11) Re(S21) Im(S21) Re(S12) Im(S12) Re(S22) Im(S22)
in the file) are assigned to the test ports that the analyzer actually calibrates as follows: Port 1 is
always assigned to the lower-numbered calibrated test port, port 2 to the other (higher-
numbered) calibrated test port.
Assigning a label to standards is optional. However, the label is displayed in many dialogs and
can provide useful information about the standard, e.g. its serial number.
If
Circuit Model
is selected in the
S-params From
panel, then the controls in the central panel of the
dialog are enabled. The circuit diagram is adjusted to the selected standard type. The following
parameters can be set:
•
Frequency range
(Min. Freq.
to
Max. Freq)
for which the circuit model is valid. During
calibration, the analyzer checks whether the sweep range is contained in the validity range of
all measured standards and possibly generates a warning (see
Measure Standards
dialog).
•
Offset
and
Load
parameters of the circuit model.
The impedance for waveguides is frequency-dependent. If a waveguide line type is selected in
the Offset Model dialog, the circuit model indicates varies instead of a definite impedance value.
Remote control:
[SENSe<Ch>:]CORRection:CKIT:<std_type>
[SENSe<Ch>:]CORRection:CKIT:<conn_type>:<std_type>
MMEMory:LOAD:CKIT:SDATa
Restrict Port Assignment
Opens a dialog to define whether the standard can be connected to any port of the analyzer or to just
one port (for one-port standards) or a pair of ports (for two-port standards).
The port assignment is displayed in the
Add or View / Modify Calibration Kit
dialogs.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......