![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 135](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759135.webp)
R&S
®
ZVA/ZVB/ZVT
Screen Elements
1145.1084.12 3.25
E-1
Examples for special points in the inverted Smith chart:
•
The magnitude of the reflection coefficient of a short circuit (Y = infinity, U = 0) is one, its phase is
–180
0
.
•
The magnitude of the reflection coefficient of an open circuit (Y = 0, I = 0) is one, its phase is
zero.
Measured Quantities and Display Formats
The analyzer allows any combination of a display format and a measured quantity. The following rules
can help to avoid inappropriate formats and find the format that is ideally suited to the measurement
task.
•
All formats are suitable for the analysis of reflection coefficients S
ii
. The formats
SWR, Smith
and
Inverted Smith
lose their original meaning (standing wave ratio, normalized impedance or
admittance) if they are used for transmission
S-parameters
,
Ratios
and other quantities.
•
The complex
Impedances, Admittances, Z-parameters,
and
Y-parameters
are generally
displayed in one of the Cartesian diagrams with linear vertical axis scale or in a polar diagram.
•
The
real
Stability Factors, DC Inputs,
and the
PAE
is generally displayed in a linear Cartesian
diagram
(Lin Mag
or
Real).
In complex formats, real numbers represent complex numbers with
zero imaginary part.
The following table gives an overview of recommended display formats.
Complex
dimensionless
quantities:
S-parameters and ratios
Complex quantities with dimensions:
Wave quantities, Z-parameters, Y-
pParameters, impedances, admittances
Real quantities:
Stability factors, DC
input 1/2, PAE
Lin Mag
(default for Z-parameters, Y-
parameters, impedances, admittances)
(default)
dB Mag
(default)
(default for wave quantities)
–
Phase
–
Real
Imag
–
Unwrapped
Phase
–
Smith
(reflection coefficients S
ii
)
– –
Polar
– –
Inverted Smith
(reflection coefficients S
ii
)
– –
SWR
(reflection coefficients S
ii
)
– –
Delay
(transmission coefficients
S
ij
)
– –
The default formats are activated automatically when the measured quantity is changed.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......