R&S
®
ZVA/ZVB/ZVT
Trace
Menu
1145.1084.12 4.25
E-6
More Admittances
Opens a dialog to select converted, matched-circuit admittances for more ports or balanced port
measurements.
The notation for admittance parameters and the functionality of the
More Admittances
dialog is
analogous to the definition of S-parameters.
•
Y<—
selects the type (left pull-down list) and the port number assignment (right pull-down list)
of the admittance parameter. Mixed mode parameters are only available if a balanced port
configuration is active. They are expressed as Y
<mout><min>
, where <mout> and <min> denote the
output and input port modes. The port numbers are assigned in the order Y
<out>< in>
, where
<out> and <in> denote the output (response) and input (stimulus) port numbers of the DUT.
•
Balanced and Measured Ports
opens the
dialog of the same name to define the properties of
the test ports. Single-ended (unbalanced) admittance parameters are assigned to the physical
test ports of the analyzer. Balanced admittance parameters are assigned to logical test ports.
Selecting a balanced port configuration with logical test ports means that the balance-
unbalance conversion is switched on and that the analyzer provides mixed mode parameters.
•
The graphics above the
Balanced and Measured Ports
button illustrates the current port
configuration.
The port configuration is valid for all traces in the channel
The settings made in the
Balanced and Measured Ports
dialog are channel settings and therefore apply
to all traces assigned to the channel
(Balanced and Measured Ports
is also accessible through
Channel
–
Mode
–
Port Config...
). Within a channel the analyzer measures either single-ended or mixed mode
parameters. If a balanced test port configuration is selected the single-ended parameters assigned to
the channel are converted into mixed mode parameters.
The relation between the generalized (multiport and mixed mode) admittance parameters
and the S-parameters is analogous to the 2-port case and follows from the relations for
impedance parameters.
Remote
control:
CALCulate<Ch>:PARameter:MEASure "<Trace_Name>", "SY11" |
"SY12" ...
Create new trace and select name and measurement parameter:
CALCulate<Ch>:PARameter:SDEFine "<Trace_Name>", "SY11" |
"SY22" | ...
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......