![Rohde & Schwarz 1145.1010.04/05/06 Operating Manual Download Page 745](http://html1.mh-extra.com/html/rohde-and-schwarz/1145-1010-04-05-06/1145-1010-04-05-06_operating-manual_1477759745.webp)
R&S
®
ZVA/ZVB/ZVT
SOURce<Ch>:POWer...
1145.1084.12 6.283
E-6
Note:
If the start power entered is greater than the current stop power
(
SOURce<Ch>:POWer<Pt>:STOP
), the stop power is set to the start power plus the minimum
power span of 0.01 dB.
SOURce<Ch>:POWer<Pt>:STATe <Boolean>
Turns the RF source power at a specified test port on or off.
<Ch>
Channel number.
<Pt>
Test port number of the analyzer.
<Boolean>
ON | OFF - Turns the internal source power at the specified test
port no. <Pt> on or off.
*RST value
ON
SCPI, Command
Types
Device-specific, command or query
Example:
*RST; SOUR:POW -6
Set the internal source power for channel 1 and all test ports to -6
dBm.
SOUR:POW2:STAT?
Query whether the source power at test port 2 is on. The
analyzer returns a 1.
SOURce<Ch>:POWer<Pt>:STOP <numeric_value>
Defines the stop power for a power sweep which is equal to the right edge of a Cartesian diagram. A
power sweep must be active (
[SENSe<Ch>:]SWEep:TYPE POWer
) before this command can be
used.
<Ch>
Channel number.
<Pt>
Test port number of the analyzer. This suffix is ignored because
the selected power sweep range applies to all source ports used
in the active channel.
<numeric_value>
Stop power of the sweep.
Range [def. unit]
–39.99 dBm to +10 dBm. The exact range depends on the
analyzer model; refer to the data sheet [dBm]. UP and DOWN
increment/decrement the source power in 1-dB steps.
*RST value
0 dBm
SCPI, Command
Types
Confirmed, command or query
Example:
SWE:TYPE POW
Activate a power sweep.
SOUR:POW:STAR -6; STOP 10
Select a power sweep range between –6 dBm and +10 dBm.
Summary of Contents for 1145.1010.04/05/06
Page 10: ......
Page 20: ......
Page 22: ......
Page 48: ......
Page 70: ......
Page 72: ......
Page 90: ......
Page 92: ......
Page 108: ......
Page 156: ......
Page 162: ......
Page 406: ...Display Menu R S ZVA ZVB ZVT 1145 1084 12 4 244 E 6 Stack Tile Horizontally Tile Vertically...
Page 450: ...Status Reporting System R S ZVA ZVB ZVT 1145 1084 12 5 18 E 1...
Page 462: ......
Page 766: ......
Page 772: ......
Page 792: ......
Page 794: ......
Page 808: ......