3-73
3.2 Messages and Actions
3
NOTE 1: The refreshing backup can be restored (see Section 3.4).
W-4386
The sliding stage is not installed.
Couldn’t perform the scanner
correction.
Install it.
Scanner correction demand is
received when the mask table slider
is not set.
Set the mask table slider and then
perform the scanner correction.
E 4387
A communication error occurred
between the main control section and
GMB circuit boards.
Restart the system. If problems
persist, consult your technical
representative.
Start response timeout
(Causes of the error message)
1. Poorly connected or broken 1394
interface cable
2. Poorly inserted GIE23 circuit
board
3. Faulty GIE23 circuit board
4. Faulty GMB23 circuit board
1. Connect the cable correctly, or
repair or replace it.
2. Insert the circuit board correctly in
the slot.
3. Replace the GIE23 circuit board.
4. Replace the GMB23 circuit board.
E-4388
Initializing of communication for
IEEE1394 failed.
Restart the system. If problems
persist, consult your technical
representative.
1394 node connection failure
(Causes of the error message)
1. Poorly connected or broken 1394
interface cable between the image
processing box and the main
control unit
2. GMB23 circuit board does not
start up.
1. Connect the cable correctly or
replace it.
2. Inspect and repair the power
supply system of the GMB23
circuit board. Replace the GMB23
or GIE23 circuit board.
E-4389
Initializing of communication for
IEEE1394 failed.
Restart the system. If problems
persist, consult your technical
representative.
1394 startup failure
Restart the system software.If the
error occurs again, reinstall the
system software. (NOTE1)
W-4390
Couldn’t generate image log file due
to insufficient disk space.
Restart the system. If problems
persist, consult your technical
representative.
Image log cannot be stored due to
lack of disk capacity.
Delete unnecessary files from the D
drive of the main control unit.
E-4391
Couldn't generate log file.
Restart the system. If problems
persist, consult your technical
representative.
Log cannot be stored due to lack of
hard disk capacity.
(Causes of the error message)
1. Lack of hard disk capacity.
1. Check the capacity of the hard
disk and delete unnecessary files
(such as templates).
E-4392
A communication error occurred
between the image processing
section and the carrier.
Restart the system. If problems
persist, consult your technical
representative.
No response obtained by command
to carrier block
Restart the system software. If the
same error occurs again, reinstall the
system software. (NOTE 1)
E-4393
A communication error occurred
between the image processing
section and scanner light source unit.
Restart the system. If problems
persist, consult your technical
representative.
No response obtained by command
to scanner light source block
Restart the system software. If the
same error occurs again, reinstall the
system software. (NOTE 1)
No.
Message
Factor
Actions