3-56
3.2 Messages and Actions
E-2659
P2RB replenishment flow volume or
its solution level abnormal.
About 510 3R-size sheets can still be
printed without replenishment.
Consult your technical representative.
Accumulated output amount of
solution to P2RB upper level sensor
exceeds the specified value (777.0
ml).
(Causes of the error message)
1. Faulty P2RB pump (PU723)
2. Faulty P2RB cartridge washing
valve (S730)
3. Faulty P2RB upper level sensor
(FS722)
4. Bent or clogged P2RB hose
1.
2.
3.
Take appropriate corrective
measures.
4. Repair or replace the hose.
I-2662
Performing evaporation correction.
Warming-up will be started after the
evaporation correction.
Please wait. ( Max 25 min.)
Performing evaporation correction.
E-2663
P1R replenishment pump(PU721) or
lower limit sensor(FS724) abnormal.
About 510 3R-size sheets can still be
printed without replenishment.
Consult your technical representative.
Accumulated output amount of
solution between P1R upper and
lower sensors exceeds the specified
value.
(Causes of the error message)
1. Faulty P1R pump (PU721)
2. Faulty P1R lower level sensor
(FS724)
3. Incorrect default setting of pump
output amount
4. Bent or clogged P1R hose
1.
2.
Take appropriate corrective
measures.
3. Perform Menu 0620 “Replenisher
Pump Output Measurement/
Setting”.
4. Repair or replace the hose.
E-2664
P2RA replenishment flow volume or
its solution level abnormal.
About 510 3R-size sheets can still be
printed without replenishment.
Consult your technical representative.
Accumulated output amount of
solution between P2RA upper and
lower sensors exceeds the specified
value.
(Causes of the error message)
1. Faulty P2RA pump (PU722)
2. Faulty P2RA lower level sensor
(FS725)
3. Incorrect default setting of pump
output amount
4. Bent or clogged P2RA hose
1.
2.
Take appropriate corrective
measures.
3. Perform Menu 0620 “Replenisher
Pump Output Measurement/
Setting”.
4. Repair or replace the hose.
E-2665
P2RB replenishment flow volume or
its solution level abnormal.
About 510 3R-size sheets can still be
printed without replenishment.
Consult your technical representative.
Accumulated output amount of
solution between P2RB upper and
lower sensors exceeds the specified
value.
(Causes of the error message)
1. Faulty P2RB pump (PU723)
2. Faulty P2RB lower level sensor
(FS726)
3. Incorrect default setting of pump
output amount
4. Bent or clogged P2RB hose
1.
2.
Take appropriate corrective
measures.
3. Perform Menu 0620 “Replenisher
Pump Output Measurement/
Setting”.
4. Repair or replace the hose.
No.
Message
Factor