3-100
3.2 Messages and Actions
NOTE 1: The refreshing backup can be restored
(See Section 3.4)
.
E-4625
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-serial Transmission Driver 3
overrun and parity error
(Causes of the error message)
1. Noise in electric line
- Outside noise (static electricity,
radio transmission or power
supply-induced)
- Inside noise
2. System software overload (faulty
software)
3. Faulty CLE23 circuit board (CPU)
1. Click [OK] to clear the message.
2. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
3. Replace the circuit board.
E-4626
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-serial Transmission Driver 3
overrun and framing error
(Causes of the error message)
1. Noise in electric line
- Outside noise (static electricity,
radio transmission or power
supply-induced)
- Inside noise
2. System software overload (faulty
software)
3. Faulty CLE23 circuit board (CPU)
1. Click [OK] to clear the message.
2. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
3. Replace the circuit board.
E-4627
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-serial Transmission Driver 3
overrun, framing and parity error
(Causes of the error message)
1. Noise in electric line
- Outside noise (static electricity,
radio transmission or power
supply-induced)
- Inside noise
2. System software overload (faulty
software)
3. Faulty CLE23 circuit board (CPU)
1. Click [OK] to clear the message.
2. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
3. Replace the circuit board.
E-4628
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-serial Transmission Driver 3
sending timeout
(Causes of the error message)
1. Noise in electric line
- Outside noise (static electricity,
radio transmission or power
supply-induced)
- Inside noise
2. System software overload (faulty
software)
3. Faulty CLE23 circuit board (CPU)
1. Click [OK] to clear the message.
2. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
3. Replace the circuit board.
E-4629
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-transmission 2 message
sending error
(Causes of the error message)
1. Noise in electric line
- Outside noise (static electricity,
radio transmission or power
supply-induced)
- Inside noise
2. System software overload (faulty
software)
3. Faulty CLE23 circuit board (CPU)
1. Click [OK] to clear the message.
2. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
3. Replace the circuit board.
No.
Message
Factor
Actions