3-96
3.2 Messages and Actions
NOTE 1: The refreshing backup can be restored
(See Section 3.4)
.
E-4592
An error occurred during accessing of
the CTB circuit board external
memory.
Restart the system. If problems
persist, consult your technical
representative.
SKCTB-EEPROM writing abnormal
(Causes of the error message)
1. Faulty system software
2. Faulty CTB23 circuit board (CPU
or EEPROM)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4595
Detecting of the shutter home
position failed.
Restart the system. If problems
persist, consult your technical
representative.
SKCTB-shutter home position
operation abnormal 3
(Causes of the error message)
1. Faulty shutter home position
sensor (D203) harness
2. Faulty shutter drive motor (M203)
harness
1. Reconnect or replace the harness.
2. Reconnect or replace the harness.
E-4596
Detecting of the shutter home
position failed.
Restart the system. If problems
persist, consult your technical
representative.
SKCTB-shutter home position
operation abnormal 3
(Causes of the error message)
1. Faulty shutter home position
sensor (D203) harness
2. Faulty shutter drive motor (M203)
harness
1. Reconnect or replace the harness.
2. Reconnect or replace the harness.
E-4597
The shutter control is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCTB-shutter opening abnormal
(Causes of the error message)
1. Faulty shutter home position
sensor (D203) harness
2. Faulty shutter drive motor (M203)
harness
1. Reconnect or replace the harness.
2. Reconnect or replace the harness.
E-4598
The shutter control is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCTB-shatter opening timeout error
(Causes of the error message)
1. Faulty shutter home position
sensor (D203) harness
2. Faulty shutter drive motor (M203)
harness
1. Reconnect or replace the harness.
2. Reconnect or replace the harness.
E-4599
The shutter control is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCTB-shutter closing abnormal
(Causes of the error message)
1. Faulty shutter home position
sensor (D203) harness
2. Faulty shutter drive motor (M203)
harness
1. Reconnect or replace the harness.
2. Reconnect or replace the harness.
E-4600
The shutter control is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCTB-shutter closing timeout error
(Causes of the error message)
1. Faulty shutter home position
sensor (D203) harness
2. Faulty shutter drive motor (M203)
harness
1. Reconnect or replace the harness.
2. Reconnect or replace the harness.
E-4601
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-message delivery module
initialization error 1
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4602
The control for the CLE circuit board
is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-message delivery module
initialization error 2
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
No.
Message
Factor
Actions