3.2 Messages and Actions
NOTE 1: The refreshing backup can be restored
(See Section 3.4)
.
W-4354
An incorrect parameter in the CPZ
circuit board.
Consult your technical
representative.
Parameter of the scanner CPZ23
circuit board differs from one of the
HDD.
(Causes of the error message)
1. CPZ23 circuit board has been
replaced.
2. Main control unit has been
replaced.
3. Parameter writing is not performed
after the system software has
been newly installed.
4. Damaged data of circuit board or
file in HDD of main control unit
HDD.
On Menu 0350 “Scanner Parameter
Check/Update”, perform the
following.
1. Click [Reading].
2. Click [Storage].
3. Click [Storage].
4. Perform “AD Timing Adjustment”,
“OFD voltage adjustment” and
“Gray Pixel Detection”.
W-4355
Writing data to hard disk failed.
Consult your technical
representative.
Disk writing failure
(Causes of the error message)
1. Lack of HDD capacity
2. Faulty software
3. Faulty HDD
1. Delete unnecessary files from the
HDD.
2. Restart the system software.If the
error occurs again, reinstall the
system software. (NOTE1)
3. Replace the HDD.
W-4356
Reading data from hard disk failed.
Consult your technical
representative.
Reading from disk failure
(Causes of the error message)
1. Faulty software
2. Faulty HDD
1. Restart the system software.If the
error occurs again, reinstall the
system software. (NOTE1)
2. Replace the HDD.
W-4357
No files are found in the hard disk.
Consult your technical
representative.
Search disk file failure
(Causes of the error message)
1. Faulty software
2. Faulty HDD
1. Restart the system software.If the
error occurs again, reinstall the
system software. (NOTE1)
2. Replace the HDD.
W-4358
Please install the calibration mask.
Couldn’t perform the scanner
correction.
Calibration mask not installed.
(Causes of the error message)
1. Calibration mask not installed.
2. Faulty mask sensor
1. Install the calibration mask.
2. Replace the MSB23 circuit board.
W-4359
Unknown carrier type.
Install the carrier again.
If problems persist, consult your
technical representative.
Incorrect carrier code is received.
(Cusses of the error message)
1. Program in carrier changed for
some reason.
2. Scanner software version
unsuitable for this carrier.
1. If the type differs between the
installed carrier and the displayed
one, replace the carrier circuit
board.
2. Upgrade the system software (A1).
W-4360
The focus position adjustment is not
performed.
Perform the focus position
adjustment with the 135 full-frame
mask.
135F mask focus position adjustment
is not performed when scanner
correction with NC100AY carrier is
requested.
Perform and register “Focus Position
Adjustment”.
W-4361
The focus position adjustment is not
performed.
Perform the focus position
adjustment with the 135 half-frame
mask.
135H mask focus position adjustment
is not performed when scanner
correction with NC100AY carrier is
requested.
Perform and register “Focus Position
Adjustment”.
W-4362
The focus position adjustment is not
performed with this mask.
Perform it with this mask.
Scanning operation is requested with
mask not performed focus position
adjustment.
Perform and register “Focus Position
Adjustment”.