5-122
5.7 Processor Adjustment/Maintenance (06)
9. PSR :
Dispose of the liquid in the measuring cylinder.
P1R, P2RA or P2RB :
Return the discharged replenisher into the
replenisher tank through the cartridge box
opening.
10. Select the “Measured value” box, and enter the
recorded output value.
11. Clean the measuring cylinder.
12. Click the [OK] button.
• Operation returns to the menu screen.
13. Reinstall the removed parts.
This procedure measures and corrects auto-cleaning
pump output.
[Selection]
[Setup and Maintenance]
→
[06 Processor
Adjustment/Maintenance]
→
[0621 Auto Cleaning
Output Measurement/Setting]
1. Prepare the two measuring cups for auto-cleaning
output measurement.
2. Open the dryer unit and support it with Stopper 2
(yellow).
3. Remove the No.2 crossover rack.
NOTE: Do not remove the other crossover racks to
prevent splashing of washing water.
L
2
1052
P1R
Replenisher
Nozzle
Measuring
Cylinder
P2RB Replenisher Nozzle
P2RA
Replenisher
Nozzle
P1W
Replenisher
Nozzle
5.7.2
Auto Cleaning Output Measurement/Setting (0621)
L
2
1052
No.2 Crossover Rack
No.3 Crossover Rack