3-118
3.2 Messages and Actions
E-4761
Electrical failure occurred.
(CCD CB -12V)
Restart the system. If problems
persist, consult your technical
representative.
(Causes of the error message)
1. Poorly connected CN2 to CCD
circuit board
2. Poorly connected JCD1 connector
to CCD unit
3. Poorly connected PS6 connector
to power supply unit
4. Broken harness
5. Blown fuse in power supply
6. Faulty CCD23 circuit board
7. Faulty power supply unit
1. Reconnect the connector.
2. Reconnect the connector.
3. Reconnect the two connectors.
4. Replace the harness.
5. Replace the fuse.
6. Replace the CCD unit.
7. Replace the power supply unit.
E-4762
Electrical failure occurred.
(CCD CB +12V)
Restart the system. If problems
persist, consult your technical
representative.
(Causes of the error message)
1. Poorly connected CN2 to CCD
circuit board
2. Poorly connected JCD1 connector
to CCD unit
3. Poorly connected PS6 connector
to power supply unit
4. Broken harness
5. Blown fuse in power supply
6. Faulty CCD23 circuit board
7. Faulty power supply unit
1. Reconnect the connector.
2. Reconnect the connector.
3. Reconnect the two connectors.
4. Replace the harness.
5. Replace the fuse.
6. Replace the CCD unit.
7. Replace the power supply unit.
W-4763
The accumulation time in auto-
exposure adjustment reaches the
maximum value.
Check if there is no preventive object
in the optical path.
At the time of scanner correction or
brightness correction creation
(Causes of the error message)
1. Foreign matter in optical path
2. Faulty harness between CCD23
and GIA23 circuit boards
3. Faulty CCD
1. Remove the foreign matter.
2. Reconnect or replace the harness.
3. Replace the CCD unit.
E-4764
The external memory of GIA circuit
board is abnormal.
Restart the system. If problems
persist, consult your technical
representative.
Result of GIA23 circuit board outside
memory R/W was NG.
(Causes of the error message)
1. Poorly inserted GIA23 circuit
board into slot of GMB23 circuit
board
2. Faulty GIA23 circuit board
1. Insert the circuit board into the slot
correctly.
2. Replace the circuit board.
E-4772
An error occurred during accessing of
the CLE circuit board external
memory.
Restart the system. If problems
persist, consult your technical
representative.
Error occurs starting up, or executing
lens registration/magnification
calibration/focus calibration/LED light
amount adjustment/scanner
parameter check.
(Causes of the error message)
1. Faulty software
2. Faulty CLE23 circuit board (CPU
or EEPROM)
1. Restart the system.
2. Replace the CLE23 circuit board.
(Make sure the system starts up
normally and the parameter can
be rewritten).
W-4801
Gaining of the image failed.
Image obtaining failure
(Causes of the error message)
1. IEEE1394 interface of GIE23
circuit board or main control unit is
unstable.
2. Faulty application on GMB23
circuit board
3. Faulty 1394 cable from image
processing box to main control unit
1. Restart the system software. If the
error occurs again, replace the
GIE23 or GMB23 circuit board.
2. Restart the system software. If the
error occurs again, replace the
GIE23 or GMB23 circuit board.
3. Reconnect or replace the cable.
W-4802
Auto focus failed.
AF was executed without chart jig or
film in mask.
Set the chart jig or film and execute
again.
I-4803
Performing the focus calibration.
At the time of focus calibration
Wait until calibration is completed.
W-4804
The lens is not registered.
Register it.
Lens was not registered at the time of
optical magnification calibration.
Register the lens.
No.
Message
Factor
Actions