3-71
3.2 Messages and Actions
3
NOTE 1: The refreshing backup can be restored (see Section 3.4).
I-4363
The scanner temperature control is
not completed.
Please wait.
After the completion of temperature
control, the selected menu will be
automatically started.
Temperature control not completed.
Wait until completion.
I-4364
The diffusion box is incorrect.
Couldn’t perform the scanner
correction.
Check carrier and diffusion box.
Diffusion box type cannot be detected
when scanner correction is
requested.
(Causes of the error message)
1. Diffusion box not installed.
2. Faulty diffusion box sensor.
3. Broken or shorted diffusion box
sensor harness
1. Install the diffusion box.
2. Replace the sensor.
3. Repair or replace the harness.
I-4365
Closing the maintenance process.
Please wait.
Maintenance completing
Wait until completion.
W-4366
The scanner correction was
cancelled.
Try it again.
Operation occurred to cancel scanner
correction
Perform the scanner correction again.
E-4369
Scanning conditions are not decided.
Check if the diffusion box, carrier or
mask is correctly installed.
If this error persists, consult your
technical representative.
Scanning condition (combination of
diffusion box, carrier and mask)
decision error
(Causes of the error message)
1. Incorrectly seated mask
2. Faulty mask sensor
1. Install the mask correctly.
2. Replace the MSB23 circuit board.
E-4370
An incorrect scanning condition was
entered.
Check the printing magnification and
the scanning range.
If this error persists, consult your
technical representative.
Entered parameter (magnification,
range etc.) is zero or below.
(Causes of the error message)
1. Incorrect print size value is set up.
2. Damaged scanning parameter in
HDD
1. Set print size up correctly.
2. Reinstall the system software (A1).
(NOTE 1)
W-4371
It is unaccepted to print using this
magnification.
Decrease the printing magnification.
Parameter is specified to increase
pixel shifting when executing fine
scanning in one-way mode.
(Causes of the error message)
1. Damaged scanning condition
parameter file in HDD
1. Restart the system software.If the
error occurs again, reinstall the
system software. (NOTE1)
W-4372
The printing magnification exceeds
the upper limit.
Decrease the printing magnification.
Printing magnification exceeds upper
limit.
(Causes of the error message)
1. Damaged scanning condition
parameter file in HDD
1. Restart the system software.If the
error occurs again, reinstall the
system software. (NOTE1)
W-4373
The printing magnification exceeds
the lower limit.
Increase the printing magnification.
Printing magnification exceeds lower
limit.
(Causes of the error message)
1. Damaged scanning condition
parameter file in HDD
1. Restart the system software.If the
error occurs again, reinstall the
system software. (NOTE1)
W-4374
The cropping range is too small.
Increase the cropping range.
Too small cropping range
Increase the cropping range.
No.
Message
Factor
Actions