3-75
3.2 Messages and Actions
3
NOTE 1: The refreshing backup can be restored (see Section 3.4).
E-4405
A time-out error occurred during the
main control section command was
issued.
Restart the system. If problems
persist, consult your technical
representative.
No response obtained when sending
command to main control unit
(Causes of the error message)
1. Poorly connected 1394 cable
2. Faulty system software
1. Connect the cable correctly.
2. Restart the system software. If the
same error occurs again, reinstall
the system software. (NOTE 1)
E-4406
A time-out error occurred during the
auto dark correction table was being
created.
Restart the system. If problems
persist, consult your technical
representative.
No response obtained when sending
auto darkness correction table
creation command to GIA 23 circuit
board
(Causes of the error message)
1. Poorly connected harness
(CCD23 to GIA23 circuit board)
2. Faulty GIA23 circuit board
3. Faulty CCD23 circuit board
1. Connect the harness correctly.
2. Replace the circuit board.
3. Replace the CCD unit.
E-4408
The downloading of the data from the
main control section failed.
Restart the system. If problems
persist, consult your technical
representative.
Download completion response is not
obtained from the main control unit or
download response from the main
control unit is NG.
(Causes of the error message)
1. Poorly connected 1394 cable
2. Faulty system software
1. Connect the cable correctly.
2. Restart the system software. If the
same error occurs again, reinstall
the system software. (NOTE 1)
E-4410
Uploading of the data to the main
control section failed.
Restart the system. If problems
persist, consult your technical
representative.
Upload response from the main
control unit is NG.
(Causes of the error message)
1. Faulty 1394 cable between GMB
circuit board and main control unit
2. Faulty HDD in main control unit
1. Reconnect or replace the 1394
cable.
2. Replace the HDD.
E-4411
Creating of the bright correction file
information failed.
Restart the system. If problems
persist, consult your technical
representative.
Brightness correction response
failure
Restart the system software. If the
same error occurs again, reinstall the
system software. (NOTE 1)
W-4412
Auto focus failed.
Check if the film is set correctly.
Peak detecting retry over at AF
execution
(Causes of the error message)
1. Film (error occurs by specific
frame or film)
• Low contrast
• Heavily curled films
• Excessively thin or thick mounts
2. Carrier (situation is changed by
carriers or masks)
• Carrier not seated on carrier
base correctly
• Mask not seated correctly
• Faulty carrier pressure system
3. Scanner (frequent cavses)
• Abnormal focus position
adjustment value
• Abnormal focus calibration value
• Loss of synchronization of lens
or conjugate length motor
1. Correct the film (replace the
mount).
2. Remove the carrier and reinstall it
or perform carrier self-diagnostics.
3. Close “I/O Check” screen to check
the mechanical initializing error
and then inspect the lens and
conjugate length variable drive
section.
No.
Message
Factor
Actions