3.2 Messages and Actions
3
W-2668
SU2400AY alignment operation
(M814/D814) abnormal.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Abnormal print alignment
(Causes of the error message)
1. Poorly connected or broken sorter
cable
2. Poorly connected or broken
harness between the print
alignment motor for SU2400AY
(M812) and PDB10 on the PDB23
circuit board
3. Poorly connected or broken
harness between the print
alignment stop sensor (D814) and
JNE4 on the JNE23 circuit board
4. Faulty M812
5. Faulty D814
6. Faulty PDB23 circuit board
7. Increased load in the drive system
for print alignment
8. Prints are hit by a sorter tray.
1. Connect the cable correctly, or
repair or replace it.
2. Connect the harness correctly, or
repair or replace it.
3. Connect the harness correctly, or
repair or replace it.
4. Replace M812.
5. Replace D814.
6. Replace the PDB23 circuit board.
7. Repair the drive system.
8. Remove the sorter and then
reinstall it.
W-2670
Replenisher cartridge (D721) not
loaded.
Install new replenisher cartridge.
Replenisher cartridge is not installed.
(Causes of the error message)
1. Replenisher cartridge is not
installed.
2. Poorly connected or broken
harness between the replenisher
cartridge setting sensor (D721)
and the JNE2 on the JNE23 circuit
board
3. Faulty D721
1. Install a new replenisher cartridge.
2. Connect the harness properly, or
repair or replace it.
3. Replace D721.
W-2671
Replenisher door (D724) open.
Close it.
Replenisher box door is open.
(Causes of the error message)
1. The replenisher box door is open.
2. Poorly connected or broken
harness between the replenisher
box door open/close detecting
interlock switch (D724) and PWR4
on the PWR23 circuit board
3. Faulty D724
4. Faulty PWR23 circuit board
1. Close the replenisher box door.
2. Connect the connector properly, or
repair or replace it.
3. Replace D724.
4. Replace the PWR23 circuit board.
W-2672
Mix the replenisher.
Add 8L of PSR.
When the PSR upper limit level
sensor detects “no solution”
(Causes of the error message)
1. PSR replenisher solution level is
low.
2. Faulty replenisher solution level
sensor system
1. Add (about 8 liters of) FRSS until
this message disappears.
2.
E-2673
Opening of replenisher cartridge
failed.
Consult your technical representative.
When a lower level sensor does not
detect “solution exists” within a
specified time after the cartridge is
opened.
(Causes of the error message)
1. Replenisher cartridge is opened
improperly.
2. Clogged replenishment hose.
3. Faulty lower level sensor system
1.
2.
3.
No.
Message
Factor
Actions