3-103
3.2 Messages and Actions
3
NOTE 1: The refreshing backup can be restored (see Section 3.4).
E-4645
An error occurred during accessing of
the CLE circuit board external
memory.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-I2C slave a R/W
data sending error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU
or EEPROM)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4646
An error occurred during accessing of
the CLE circuit board external
memory.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-I2C slave a R/W
data sending failure
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU
or EEPROM)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4647
An error occurred during accessing of
the CLE circuit board external
memory.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-I2C higher-order address
sending error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU
or EEPROM)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4648
An error occurred during accessing of
the CLE circuit board external
memory.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-I2C higher-order address
sending failure
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU
or EEPROM)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4549
An error occurred during accessing of
the CLE circuit board external
memory.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-I2C lower-order address
sending error
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU
or EEPROM)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4650
An error occurred during accessing of
the CLE circuit board external
memory.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-I2C lower-order address
sending failure
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU
or EEPROM)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4651
An error occurred during accessing of
the CLE circuit board external
memory.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-I2C start bit sending error 2
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU
or EEPROM)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
E-4652
An error occurred during accessing of
the CLE circuit board external
memory.
Restart the system. If problems
persist, consult your technical
representative.
SKCLE-I2C slave a R/W
data sending error 2
(Causes of the error message)
1. Faulty system software
2. Faulty CLE23 circuit board (CPU
or EEPROM)
1. Restart the system software. If the
error occurs again, reinstall the
system software. (NOTE 1)
2. Replace the circuit board.
No.
Message
Factor
Actions