3-78
3.2 Messages and Actions
W-4426
Detecting of the mask failed.
Install the carrier and the mask
correctly.
At scanner correction, brightness
correction or defect pixel creating
(Causes of the error message)
1. Required 135 upper and lower
mask not installed
2. Diffusion plate of 135 lower mask
removed
3. Foreign matter in scanner opening
4. Scanner optical axis misalignment
1. Install the required 135 mask.
2. Fit the diffusion plate to the 135
lower mask or replace the mask.
3. Remove the foreign matter.
4. Adjust the optical axis.
E-4429
The conjugate length control is
abnormal.
Restart the system. If problems
persist, consult your technical
representative.
At post-operational checks
(Causes of the error message)
1. Faulty harness
2. Overload (foreign matter, lack of
lubricant)
3. Faulty CTB23 circuit board or
power supply
1. Replace the harness.
2. Remove foreign matter or apply
grease.
3. Replace the circuit board or power
supply unit.
E-4430
The lens is stepped out.
Restart the system. If problems
persist, consult your technical
representative.
At post-operational checks
(Causes of the error message)
1. Faulty harness
2. Overload (foreign matter, lack of
lubricant)
3. Faulty CTB23 circuit board or
power supply
1. Replace the harness.
2. Remove foreign matter or apply
grease.
3. Replace the circuit board or power
supply unit.
W-4431
The LED temperature is not
stabilized.
If this error persists, consult your
technical representative.
At one-way or two-way scanning
mode
(Causes of the error message)
1. Increased LED temperature due to
too high a room temperature (32˚C
or higher)
2. Blocked opening for light source
section exhaust fan
3. Clogged scanner air filter
4. Too low a room temperature(10˚C
or lower)
5. Faulty LED heater or LED light
source temperature sensor
6. Faulty LED heater or LED light
source temperature sensor
harness
7. Faulty CLE23 circuit board
1. Decrease the room temperature.
2. Clear the opening.
3. Clean the air filter.
4. Increase the room temperature.
5. Replace the LED heater and
sensor assembly.
6. Replace the harness.
7. Replace the circuit board.
E-4432
The version of the CTB circuit board
software is mismatched.
Consult your technical
representative.
Abnormal software at startup
initializing
Load the CTB23 circuit board
software using the system software
(A1) CD.
E-4433
The version of the CLE circuit board
software is mismatched.
Consult your technical
representative.
Abnormal software at startup
initializing
Load the CLE23 circuit board
software using the system software
(A1) CD.
E-4434
The version of the CPZ circuit board
software is mismatched.
Consult your technical
representative.
Abnormal software at startup
initializing
Load the CPZ23 circuit board
software using the system software
(A1) CD.
E-4435
All channels for IR-LED are
disconnected.
Consult your technical
representative.
Startup initializing
(Causes of the error message)
1. Faulty harness between CLE23
and LED23 circuit boards
2. Faulty LED23 circuit board
3. Faulty CLE23 circuit board
1. Reconnect, repair or replace the
harness.
2. Replace the circuit board.
3. Replace the circuit board.
No.
Message
Factor
Actions