3-41
3.2 Messages and Actions
3
E-2510
Abnormal G laser (G-SHG)
information.
The laser temperature control
stopped.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Abnormal data read from EEPROM in
the laser unit at start-up.
(Causes of the error message)
1. Poorly connected or broken
harness between the laser optical
unit and LDD13 on the LDD23
circuit board
2. Faulty LDD23 circuit board
3. Faulty laser optical unit
1. Connect the harness correctly, or
repair or replace it.
2. Replace the LDD23 circuit board.
3. Replace the laser optical unit.
E-2511
Abnormal B laser (B-LD) information.
The laser temperature control
stopped.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Abnormal data read from EEPROM in
the laser unit at start-up.
(Causes of the error message)
1. Poorly connected or broken
harness between the laser optical
unit and LDD9 on the LDD23
circuit board
2. Faulty LDD23 circuit board
3. Faulty laser optical unit
1. Connect the harness correctly, or
repair or replace it.
2. Replace the LDD23 circuit board.
3. Replace the laser optical unit.
E-2512
R laser (R-LD) temperature
adjustment outside specified range.
The laser temperature control
stopped.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Laser temperature did not converge
within a specified range during
initialization.
(Causes of the error message)
1. Poorly connected or broken
harness between the laser optical
unit and LDD11 on the LDD23
circuit board
2. Faulty LDD23 circuit board
3. Faulty laser optical unit
1. Connect the harness correctly, or
repair or replace it.
2. Replace the LDD23 circuit board.
3. Replace the laser optical unit.
E-2513
G laser (G-SHG) temperature
adjustment outside specified range.
The laser temperature control
stopped.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Laser temperature did not converge
within a specified range during
initialization.
(Causes of the error message)
1. Poorly connected or broken
harness between the laser optical
unit and LDD12/LDD14 on the
LDD23 circuit board
2. Faulty LDD23 circuit board
3. Faulty laser optical unit
1. Connect the harness correctly, or
repair or replace it.
2. Replace the LDD23 circuit board.
3. Replace the laser optical unit.
E-2514
B laser (B-LD) temperature
adjustment outside specified range.
The laser temperature control
stopped.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Laser temperature did not converge
within a specified range during
initialization.
(Causes of the error message)
1. Poorly connected or broken
harness between the laser optical
unit and LDD9 on the LDD23
circuit board
2. Faulty LDD23 circuit board
3.Faulty laser optical unit
1. Connect the harness correctly, or
repair or replace it.
2. Replace the LDD23 circuit board.
3. Replace the laser optical unit.
E-2515
Abnormal R laser (R-LD)
temperature.
The laser temperature control
stopped.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
1. Laser temperature exceeds the
range.
2. Thermistor error is detected.
(Causes of the error message)
1. Poorly connected or broken
harness between the laser optical
unit and LDD11 on the LDD23
circuit board
2. Faulty LDD23 circuit board
3. Faulty laser optical unit
1. Connect the harness correctly, or
repair or replace it.
2. Replace the LDD23 circuit board.
3. Replace the laser optical unit.
No.
Message
Factor
Actions