3.2 Messages and Actions
3
W-2650
Waste solution tank nearly full.
Collect the solution in the recovery
tank in the W2 manner.
Internal waste solution sensor
detects “Solution exists”.
(Causes of the error message)
1. Waste solution tank is full.
2. Poorly connected or broken
harness between the waste
solution level sensor (FS728) and
JNE6 on the JNE23 circuit board
3. Dirty FS728
4. Faulty FS728
1. Collect waste solution
2. Connect the harness correctly, or
repair or replace it.
3. Clean FS728.
4. Replace FS728.
W-2651
Waste solution tank full.
Printing stopped.
Collect the solution in the recovery
tank in the W2 manner.
Internal waste solution sensor
detects “Solution full”.
(Causes of the error message)
1. Waste solution tank is full.
2. Poorly connected or broken
harness between the waste
solution level sensor (FS728) and
JNE6 on the JNE23 circuit board
3. Dirty FS728
4. Faulty FS728
1. Collect waste solution.
2. Connect the harness correctly, or
repair or replace it.
3. Clean FS728.
4. Replace FS728.
E-2653
Remaining replenisher abnormal.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Status of P1R, P2RA, or P2RB level
sensor changed during standby.
(Causes of the error message)
1. Solution leakage from replenisher
tank
2. Replenisher collected or supplied
during stand-by.
3. Poorly connected or broken
harness between the solution level
sensor (FS720/FS721/FS722/
FS724/FS725/FS726) and JNE6
on the JNE23 circuit board
4. Faulty solution level sensor
(FS720/FS721/FS722/FS724/
FS725/FS726)
1. Repair solution leakage.
2. Take appropriate corrective
measures.
3. Connect the harness correctly, or
repair or replace it.
4. Replace FS720/FS721/FS722/
FS724/FS725/FS726.
E-2657
P1R replenishment flow volume or its
solution level abnormal.
About 510 3R-size sheets can still be
printed without replenishment.
Consult your technical representative.
Accumulated output amount of
solution to P1R upper level sensor
exceeds the specified value (1998.0
ml).
(Causes of the error message)
1. Faulty P1R pump (PU721)
2. Faulty P1R stirring valve (S731)
3. Faulty P1R cartridge washing
valve (S728)
4. Faulty P1R upper level sensor
(FS720)
5. Bent or clogged P1R hose
1.
2.
3.
4.
Take appropriate corrective
measures.
5. Repair or replace the hose.
E-2658
P2RA replenishment flow volume or
its solution level abnormal.
About 510 3R-size sheets can still be
printed without replenishment.
Consult your technical representative.
Accumulated output amount of
solution to P2RA upper level sensor
exceeds the specified value (777.0
ml).
(Causes of the error message)
1. Faulty P2RA pump (PU722)
2. Faulty P2RA cartridge washing
valve (S729)
3. Faulty P2RA upper level sensor
(FS721)
4. Bent or clogged P2RA hose
1.
2.
3.
Take appropriate corrective
measures.
4. Repair or replace the hose.