3-19
3.2 Messages and Actions
3
NOTE 1: The refreshing backup can be restored (see Section 3.4).
W-1912
Carrier upper cover opened.
Close it and then start pre-scan.
Scanning was attempted while the
upper cover was open.
(Causes of the error message)
1. Open pressure cover
2. Faulty pressure cover lock
mechanism
3. Faulty pressure cover open/close
sensor (D128) or poorly connected
harness
4. Poorly connected flexible cable
5. Faulty JNC23 circuit board
6. Faulty CYA23 circuit board
1. Close the pressure cover.
2. Replace the damaged parts of the
lock mechanism.
3. Replace the sensor, reconnect or
replace the harness.
4. Reconnect the flexible cable.
5. Replace the circuit board.
6. Replace the circuit board.
W-1917
Paper width and print size of negative
sheet index mismatched.
Press [OK] and replace with the
magazine of #### mm.
If you press [Output], an incorrect
negative sheet index printing will be
outputted.
This message appears when a paper
width mismatch is detected in the
format check in negative sheet index
printing.
Replace the magazine with the
correct one.
W-1918
No frame for negative sheet index
printing.
Printing of negative sheet index
cancelled.
This message appears when the
desired frame in negative sheet index
printing is not found.
–
W-1920
One-way mode cannot be used for
this carrier.
Install correct carrier.
Incorrect mask is installed for one-
way mode.
Install the correct carrier.
I-1921
You cannot return to the main menu
screen during processing of the
image.
After it is completed, return to the
main menu screen.
Printing operation is terminated
during image processing.
Wait until processing is completed.
W-1922
When the format type is ####, the
current paper width and the print size
selected as index format is
mismatched.
Press the [OK] button and replace the
paper magazine, or change the index
format.
When it is ####, the correct index
print will not be outputted after the
[Output] button is pressed.
Paper width mismatch discovered by
index format consistency check
before one-way mode pre-scanning
completion.
Replace the paper magazine or
select the index format which meets
the magazine.
W-1923
MFC mask not installed.
Install mask and perform pre-scan.
MFC10AY mask table slider is not set
properly.
Set the mask table slider correctly.
W-1924
Image data full.
Please wait a moment, and start
printing.
Print/file output processing cannot be
executed because the number of
images exceeds concurrent
processing quantity.
Wait a moment.
E-1998
Error table item(####) cannot be
omitted.
Sub-system: ####
(####)
Error No.: ####
The item(s) required for the error
table file is(are) not written.
(Causes of the error message)
1. Faulty system software
2. Faulty main control unit
1. Restart the system software.
Reinstall it if the problem occurs
again. (NOTE 1)
2. Replace the main control unit.