3-51
3.2 Messages and Actions
3
W-2609
The dryer unit (D762A) or the dryer
cover (D762B) opened.
Processing cancelled. Paper may
remain.
Press [Help] button to see error help
(or manual). Remove paper while
referring to it.
Processing tank’s or accumulating
section’s cover is opened during
paper processing.
(Causes of the error message)
1. Dryer unit or dryer belt unit is
opened.
2. Interlock switch is activated due to
vibration, etc. because the dryer
unit or dryer belt unit is installed
improperly.
3. Poorly connected or broken
harness between the dryer section
open/close detecting interlock
switch (D762) and PWR4 on the
PWR24 circuit board
4. Faulty D762
1. Close the dryer unit or dryer belt
unit.
2. Close the dryer unit or dryer belt
unit.
3. Connect the harness correctly, or
repair or replace it.
4. Replace D762.
W-2611
The dryer unit (D762A) or the dryer
section cover (D762B) opened.
Close the cover.
Processing tank’s or dryer section’s
cover is open when starting
processing.
(Causes of the error message)
1. Dryer unit or dryer belt unit is
opened.
2. Interlock switch is activated due to
vibration, etc. because the dryer
unit or dryer belt unit is installed
improperly.
3. Poorly connected or broken
harness between the dryer section
open/close detecting interlock
switch (D762) and PWR4 on the
PWR24 circuit board
4. Faulty D762
1. Close the dryer unit or dryer belt
unit.
2. Close the dryer unit or dryer belt
unit.
3. Connect the harness correctly, or
repair or replace it.
4. Replace D762.
E-2614
P1 processing tank temperature not
rising.
The heaters (H700 to H703) turned
off.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Solution temperature does not rise
more than 1 °C within 30 minutes.
(Causes of the error message)
1. Abnormal heater system
2. Abnormal temperature sensor
system
3. Abnormal circulation system
E-2615
P2 processing tank temperature not
rising.
The heaters (H700 to H703) turned
off.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Solution temperature does not rise
more than 1 °C within 30 minutes.
(Causes of the error message)
1. Abnormal heater system
2. Abnormal temperature sensor
system
3. Abnormal circulation system
E-2616
PS1/2/3 processing tank
temperatures not rising.
The heaters (H700 to H703) turned
off.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Solution temperature does not rise
more than 1 °C within 30 minutes.
(Causes of the error message)
1. Abnormal heater system
2. Abnormal temperature sensor
system
3. Abnormal circulation system
E-2617
PS4 processing tank temperatures
not rising.
The heaters (H700 to H703) turned
off.
Shut down scanner with emergency
stop button and restart it. If problems
persist, consult your technical
representative.
Solution temperature does not rise
more than 1 °C within 30 minutes.
(Causes of the error message)
1. Abnormal heater system
2. Abnormal temperature sensor
system
3. Abnormal circulation system
No.
Message
Factor
Actions