5-59
5.4 Scanner Adjustment/Maintenance (03)
5
After replacing the LED circuit board assembly or light
source unit, perform Menu 0349 “LED Light Amount
Adjustment” and then this Menu 0348 for reversal
and negative films.
[Selection]
[Setup and Maintenance]
→
[03 Scanner Adjustment/
Maintenance]
→
[0348 Spectral Calibration]
Perform the spectral calibration using the following
tools.
Tools: Chart/floppy disk set
Multi-film carrier MFC10AY
Calibration mask
135M mask
1. Insert the floppy disk into the FD drive of the main
control unit.
2. Click the [OK] button.
3. Install the Multi-film carrier MFC10AY with the
calibration mask.
4. Click the [OK] button.
5.4.14
Spectral Calibration (0348)
z
2
074
Multi-film Carrier
MFC10AY
Upper Cover
Calibration Mask