15-5
15.1 Sub-tank Section
15
3. Remove the screw and then the solution
thermometer.
Installation
Installation is essentially in the reverse order of removal.
Removal
1. Remove:
• Upper connector cover
• PS1/PS2/PS3: Processor inner cover
2. Open the five clamps and disconnect the sensor
connector.
3. Remove the screw and then the sensor.
Installation
Installation is essentially in the reverse order of removal.
LII646
Screw
Solution Thermometer (TS700)
15.1.5
Solution Level Sensor (FS700 to FS705) Replacement
LII647
Clamps (5)
Sensor Connector
(FS705)
Sensor Connector (FS704)
Sensor Connector (FS703)
Sensor Connector (FS702)
Sensor Connector (FS701)
Sensor Connector (FS700)
LII648
Screw
Sensor (FS700)