3-133
3.2 Messages and Actions
3
W-7018
An IX240 film is inserted in the lane
other than IX240 lane.
Remove the carrier and check the
position of the lane.
IX240 cartridge was inserted when
carrier was not in IX240 feed lane.
(Causes of the error message)
1. Inserted IX240 cartridge when
carrier was in another feed lane
2. Faulty cartridge set sensor (D123)
or harness
3. Faulty feed lane position sensor
(D129/D130) or harness
1. Change to the IX240 feed lane.
2. Reconnect or replace the harness,
or replace the sensor.
3. Reconnect or replace the harness,
or replace the sensor.
E-7019
Reading FlashROM data from carrier
failed.
Consult your technical
representative.
Abnormal carrier/carrier FlashROM
data
(Causes of the error message)
1. Poorly connected plug-in
connector
2. Faulty CYA23 circuit board
1. Remove and reinstall the carrier.
2. Replace the circuit board.
E-7020
Reading FlashROM data from carrier
failed.
Consult your technical
representative.
Abnormal carrier/carrier FlashROM
data
(Causes of the error message)
1. Poorly connected plug-in
connector
2. Faulty CYA23 circuit board
1. Remove and reinstall the carrier.
2. Replace the circuit board.
E-7023
Initializing of carrier failed.
Install the carrier again.
If problems persist, consult your
technical representative.
Feed roller rotation sensor (D135) did
not change when initializing.
(Causes of the error message)
1. Loose IX240 drive belt
2. Faulty feed roller rotation sensor
(D135)
3. Broken IX240 drive belt
4. Faulty feed motor(M101)
5. Faulty feed motor harness
6. Faulty CYA23 circuit board
1. Adjust belt tension.
2. Replace the sensor.
3. Replace the belt.
4. Replace the motor.
5. Reconnect or replace the harness.
6. Replace the circuit board.
W-7024
Carrier cover is opened.
Remove the carrier, and then close
carrier cover.
(Causes of the error message)
1. Open pressure cover
2. Faulty pressure cover open/close
sensor (D128)
3. Faulty harness
1. Close the pressure cover.
2. Replace the sensor.
3. Reconnect or replace the harness.
E-7025
Writing FlashROM data into carrier
failed.
Consult your technical
representative.
Writing to FlashROM failure
(Causes of the error message)
1. Poorly connected plug-in
connector
2. Faulty CYA23 circuit board
1. Remove and reinstall the carrier.
2. Replace the circuit board.