3.2 Messages and Actions
NOTE 1: The refreshing backup can be restored
(See Section 3.4)
.
W-1492
Carrier not installed.
Install carrier and press [OK].
Carrier is not installed.
(Causes of the error message)
1. Carrier is not installed.
2. Faulty carrier I/O or electric system
1. Install the carrier.
2.
W-1495
Select the correct magazine paper
type.
Press [Output] to output forcibly.
Paper types (emulsion types or
surface types) different between
upper and lower magazines
regardless of selecting “Auto” for
“Switch magazine”.
Replace magazine so that upper and
lower paper types are same.
I-1501
Selected function cannot be used.
This message appears when
selecting an unsupported function.
–
I-1502
Insert the floppy disk into the FD
drive.
This message prompts you to insert
FD during the data backup.
–
W-1503
#####
Insert the floppy disk into the FD
drive.
This message appears when FD is
not inserted during the data backup.
(Causes of the error message)
1. Floppy disk not set.
2. Floppy disk drive malfunction
1. Insert floppy disk into drive.
2. Replace main control unit.
I-1504
Insert the next floppy disk into the FD
drive.
This message prompts you to insert
the next FD during the data backup.
–
W-1505
#####
Insert the correct floppy disk into the
FD drive.
This message appears when the
inserted FD is incorrect.
(Causes of the error message)
1. Floppy disk abnormal
2. Floppy drive malfunction
1. Insert correct floppy disk into drive.
2. Replace main control unit.
E-1506
Reading of control information file
failed.
Couldn’t display the selected screen.
To use this function, consult your
technical representative.
This message appears when the files
required to display image cannot be
read.
(Causes of the error message)
1. Write protected FD
2. System software failure
3. Main control unit failure
1. Set the FD to the write enable
mode.
2. Restart system, and if problem
persists, reinstall system software.
(NOTE 1)
3. Replace the main control unit.
E-1507
Data backup completed abnormally.
This message appears when data
backup fails.
(Causes of the error message)
1. Faulty FD
2. Write protected FD
3. Lack of HDD capacity
4. System software failure
1. Replace the FD.
2. Set the FD to write enable mode.
3. Delete unnecessary files in the
HDD.
4. Reinstall the system software.
I-1508
Cancel data backup.
The data backup processing is
cancelled according to operator’s
instruction.
Click the [Cancel] button to stop the
backup.
W-1510
Date value exceeds the limit.
(1998/01/01 00:00:00 - 2037/12/31
23:59:59)
Set the correct date.
Set date value exceeds the limit.
(Causes of the error message)
1. Current date/time entry error
2. System software failure
3. Main control unit failure
1. Enter correct date/time.
2. Restart system, and if problem
persists, reinstall system software.
(NOTE 1)
3. Replace main control unit.
W-1511
Next timer-ON time exceeds the limit.
(1998/01/01 00:00:00 - 2037/12/31
23:59:59)
Confirm the present time.
Next timer-ON time exceeds the limit.
Set time value must be changed.
(Causes of the error message)
1. Current date/time entry error
2. System software failure
3. Main control unit failure
1. Enter correct date/time.
2. Restart system, and if problem
persists, reinstall system software.
(NOTE 1)
3. Replace main control unit.
No.
Message
Factor