3-22
3.2 Messages and Actions
NOTE 1: The refreshing backup can be restored
(See Section 3.4)
.
I-2305
The data in the printer backup
memory was updated.
This message does not mean
abnormality when the software
version is updated.
For other than the version up, consult
your technical representative.
–
–
E-2306
Printer's timer abnormal. Scanner
time adopted.
If problems persist, consult your
technical representative.
Error occurred in reading the time
data from timer IC at the start-up.
(Causes of the error message)
1. Abnormal system software
2. Faulty backup circuit
3. Timer IC failure
4. Memory failure
1. Restart system, and if problem
persists, reinstall system software.
(NOTE 1)
2. Replace the GPR23 circuit board.
3. Replace the GPR23 circuit board.
4. Replace the GPR23 circuit board.
I-2309
Initializing printer.
Please Wait.
While initializing the printer/
processor, the post-operational check
is activated.
–
I-2310
Printer is processing.
Please Wait.
While driving the printer/processor,
the post-operational check is
activated.
–
I-2311
Power switch in printer is turned ON.
Turn the power switch to STANDBY.
The post-operational check is
activated with the power switch
turned ON.
(Causes of the error message)
1. The power switch is ON.
1. Turn the power switch to
STANDBY.
E-2312
Timer in printer abnormal, so it was
initialized.
If problems persist, consult your
technical representative.
During system startup, abnormality of
the RTC device has been detected.
(The default value “1980/01/01” is
set.)
(Causes of the error message)
1. GPR backup switch is OFF.
2. Discharged backup battery
3. Faulty GPR circuit board
1. Turn the GPR backup switch to
ON.
2. Replace the GPR circuit board
(After starting up the system, allow
several hours for battery charging.)
3. Replace the GPR circuit board
(After starting up the system, allow
several hours for battery charging.)
I-2313
Backup memory in printer was
initialized.
During system startup, backup SRAM
clear has been detected. (Finely
adjusted values are returned to the
default settings.)
(Causes of the error message)
1. System startup after version
upgrade
2. GPR backup switch is OFF.
3. Discharged backup battery
4. Faulty GPR circuit board
1. No corrective measure is required.
2. Turn the GPR backup switch to
ON.
3. Replace the GPR circuit board
(After starting up the system, allow
several hours for battery charging.)
4. Replace the GPR circuit board
(After starting up the system, allow
several hours for charging the
battery charging.)
No.
Message
Factor
Actions