3-181
3.3 Trouble Shooting
3
3.3.13
Replenishment Cartridge Opening Malfunction
No
Yes
No
Yes
No
Yes
No
Yes
Remove replenisher cartridge.
Solution remained in cartridge due to
opening malfunction?
Any object remained in opening
nozzle hopper section?
Opening parts damaged or abnormal?
Replace parts.
Remove object.
Hose between hopper section
and tank clogged?
Correct clogging.
To “3.3.12 Return Action after
Replenishment System Error”
To “3.3.12 Return Action after
Replenishment System Error”
To “Replenishment Tank Lower Level Sensor