3-76
3.2 Messages and Actions
W-4413
Auto focus failed.
Check if the film is set correctly.
No contrast at AF
(Causes of the error message)
1. Abnormal focus position
adjustment value
2. Low contrast film
3. Scanned without film in MFC10AY
1. Perform the focus position
adjustment.
2. -
3. Insert a film and execute the
operation again.
E-4414
Auto focus failed.
Restart the system. If problems
persist, consult your technical
representative.
Motor error at AF (AF error received
from CTB23 circuit board)
(Causes of the error message)
1. Faulty harness between GIA23
and CTB23 circuit board
2. Faulty harness between CCD23
and GIA23 circuit board
1. Reconnect, repair or replace the
harness.
2. Reconnect, repair or replace the
harness.
E-4415
A time-out error occurred when the
GIA circuit board is started up.
Restart the system. If problems
persist, consult your technical
representative.
No action from GIA23 circuit board at
starting up
(Causes of the error message)
1. Poorly connected CN3 connector
to GIA23 circuit board.
2. Poorly connected connector to
CCD23 circuit board
3. Poorly connected GIA23 circuit
board in PCI slot
4. Faulty cable between GIA23 and
CCD23 circuit boards
5. Faulty GIA23 circuit board
6. Faulty CCD23 circuit board
1. Connect the connector correctly.
2. Connect the connector correctly.
3. Insert the circuit board securely.
4. Replace the cable.
5. Replace the circuit board.
6. Replace the CCD unit.
E-4417
An error occurred when the
piezoelectric actuator is controlled.
It is possible to continue printing, but
the resolution may be deteriorated
when you print with larger sizes than
4R.
Consult your technical
representative.
At the time of N-scanning fine
scanning
(Causes of the error message)
1. Temporary disturbance at
scanning (impact/high-power radio
waves)
2. Faulty CCD unit (pixel shifting
section)
3. Poorly connected or broken cable
between CCD unit and CPZ23
circuit board.
4. Faulty CPZ23 circuit board
5. Abnormal CPZ23 parameter
1. Remove disturbance.
2. Replace the CCD unit.
3. Replace the cable.
4. Replace the circuit board.
5. Check consistency of the CPZ23
circuit board parameter in
“Scanner Parameter Check/
Update”.
E-4418
The AD timing adjustment failed.
Restart the system. If problems
persist, consult your technical
representative.
FL level AD timing decision failure
(Causes of the error message)
1. Faulty CCD23 circuit board power
supply on harness
2. Faulty GIA23 circuit board power
supply harness
3. Faulty coaxial harness between
CCD23 and GIA23 circuit boards
4. Faulty serial transmission cable
between CCD23 and GIA23 circuit
board
5. Faulty CCD23 circuit board
6. Faulty GIA23 circuit board
1. Reconnect, repair or replace the
harness.
2. Replace the circuit board, and
reconnect, repair or replace the
harness.
3. Reconnect or replace the harness.
4. Reconnect or replace the cable.
5. Replace the CCD unit.
6. Replace the circuit board.
No.
Message
Factor
Actions