Repair Procedures
12-56
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
9.
Remove the belt from the motor shaft.
The motor can now be removed:
10.
Use a 3 mm Allen key to unscrew the two
screws holding the motor.
11.
Install the new motor by reversing steps 1
to 10.
Note
. When reassembling the line source compartment, route the encoder cable carefully to
prevent interference with the line source position indicator arm (see Figure 12.14 on page 12-17).
12.5.3.6
Replacing the Solenoid
The procedure for replacing the solenoid is analogous to exchanging the AC servo motor. You
perform similar steps but working in the other side of the line source compartment.
Procedure Details
Prepare for replacing the solenoid as follows:
1.
Dismount the transmission scanner from
the detector (see section 12.5.1.1, page 12-
29).
2.
Detach the line source compartment from
the bracket (see section 12.5.1.2, page 12-
33).
3.
Remove line source compartment covers
(see page 12-41).
4.
Detach the connecting cable (see section
12.5.3.2, page 12-43).
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......