NEMA Spatial Resolution and Linearity
10-10
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
6.
Replace the collimator on the selected
detector with the NEMA X line phantom
(slide the collimator out and slide in the
phantom).
The X phantom has slits in the short direction.
7.
Place the point source holder at a mini-
mum distance of 2 meters from the detec-
tor.
8.
Place a ~ 4–5 mCi (150–200 MBq) Tc-
99m point source yielding 20 kcps or less
in the holder.
Note
. Make sure that the position of the source holder is aligned with the center of the detector
FOV.
Procedure Details
9.
Make sure that the entire detector FOV is
exposed to the source and that the count
rate is < 20 kcps.
10.
Check that the phantom slits are seen as
vertical lines
11.
Click Start.
When the test is complete, the pop-up shown
to the right appears.
Note
. Do
not
click OK yet. Mount the Y slit
phantom first.
12.
Replace the NEMA X line phantom with
the Y line phantom.
The Y phantom has slits in the long direction.
13.
Click OK.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......