Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-85
Procedure Details
The Open dialog box appears.
5.
In the Look in drop-down, select the folder
C:\cardiocam
.
6.
In the File of type drop-down, select All
Files.
A listing of all the files contained in the folder
appears.
7.
Select the file
cardiocam
and click Open.
The contents of the
cardiocam
file look
similar to the example shown on the right.
8.
Find the entry
#!EnglishUnits
and edit
it as follows:
#!EnglishUnits
configures the patient
database to accept lbs., feet and inches.
#!-EnglishUnits
configures the patient
database to use metric units.
Note
. Do
not
edit any other entry of this file.
9.
Select
File
,
Save
to save the modified file.
10.
Select
File
,
Exit
to close down the
Notepad program.
11.
Double-click the CardioMD icon on the
desktop to start the CardioMD application.
To check that the correct units are being used
in the patient database:
12.
Create a new patient entry. Follow
instructions in the CardioMD Operator’s
Manual Chapter 4
Operating Instructions
.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......