Repair Procedures
12-72
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
1.
Check that the line source attenuator is in
the closed position.
If required, see instructions in section 12.5.3.9,
page 12-75 on changing the position of the
attenuator.
Attenuator in closed position
Attenuator in open position
The line source housing must be mounted
inside the line source compartment on top of
two slide bearings running on the rails in the
bottom of the compartment.
2.
Place the line source housing inside the
compartment as shown.
3.
Carefully turn the line source housing in
the direction indicated on the photo so that
it ends up being perpendicular to the rails
in the line source compartments.
Caution
. It is important to turn the line source housing in the indicated
direction when installing it. The reason for this is that the shutter lever on the
line source housing must fit into the groove in the locking plate.
The functioning of the shutter mechanism is explained on page 12-19.
CAUTION
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......