CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-87
Procedure Details
20.
Tighten the torx screws in both sides of
the AC option bracket.
21.
Remove the transportation screws from
the line source housing and check that the
line source shutter closes.
22.
Acquire a new reference scan.
See the CardioMD Operator’s Manual,
Chapter 12
CardioMD AC Option
for
instructions.
23.
Install the line source compartment covers.
Instructions are provided in section 12.5.3.10,
page 12-77.
Note
. Acquiring a new reference scan after aligning one or both transmission scanners is
mandatory.