Planned Maintenance Procedures
11-4
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
•
Verify that the rotational motion axis does not enter into any of the ultimate limits.
If an ultimate limit is reached, refer to Chapter 4
Calibration
for procedures for calibration of
absolute encoders and ultimate limits.
Also verify that the detectors are automatically stopped before hitting the patient table when driv-
ing the detectors vertically and horizontally towards the patient table.
If it is possible to drive the detectors into the patient table, refer to Chapter 4
Calibration
for pro-
cedures to calibrate the absolute position encoders.
11.3.4
Lubrication
11.3.4.1
Tools Required
•
Grease gun with lithium-based grease.
11.3.4.2
Horizontal Drive
Procedure Details
1.
Remove the cover from the gantry base to
access the horizontal drive.
2.
Apply lithium grease with the grease gun
on the zerk fitting shown in the photo to
the right.
3.
Also apply a small amount of grease to the
rails inside the gantry base and to the two
gear wheels at the motor.
11.3.4.3
Vertical Drive
Procedure Details
1.
Remove the cover on the side of the Y-
pillar.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......