1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-35
Procedure Details
3.
Remove the plate and the handle.
Note
. Note that the handle also forms part of
the coding of the transmission scanner for
detector 1 or detector 2. The handle shown in
the photo is coded for detector 2. The handle
can be mounted onto the brass disc using the
other hole to code the handle for detector 1.
4.
Turn the bracket and unscrew the cable to
the line source compartment and the cable
to the detector.
5.
Lift out the inner bracket assembly to
access the bracket components.