Detailed Description
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
3-31
Y+
Y-
DYN
COR
THRES
HOLD
DYN
COR
THRES
HOLD
DYN
COR
THRES
HOLD
DYN
COR
THRES
HOLD
DYN
COR
THRES
HOLD
Figure 3.24 Schematic of the Y-board
3.3.6.4
Tune Board
The tune board carries three octal 10-bit DAC’s. On the basis of the tuning tables, the DAC’s
provide the reference voltages for the autotune circuits in the 24 PMT’s and the control voltages
for the DynCor and threshold multipliers.
The tune board also carries a 24-to-one multiplexer and a D/A converter, enabling digital read-out
from the working points in the 24 PMT’s.
3.3.6.5
Power Board
The power board distributes high and low voltages to the six PMT interconnecting strips. It also
carries driver circuits for the autotune.
3.4
Acquisition PC
3.4.1
Tower Acquisition PC
The acquisition PC of CardoMD Series I and II systems is a tower PC with the following
minimum specifications:
IBM compatible, 550 MHz, ATX, 512 kb CPU RAM, 128 Mb SDRAM PC100, 15 Gb Hard Disk,
3.5” Floppy drive, 40x speed CD-ROM drive, 16 Mb Graphics (Matrox G400), and EtherNet
adapter (3COM XL PCI 10/100BTX).
The PC has an IEEE 1394 FireWire interface.
The PC software operates under Microsoft Windows 2000.
Newer PC’s are equipped with a CD R/W drive. The CD R/W drive can be used for backing up
calibration data and user-defined acquisition protocols.
I, II
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......