Installation
8-44
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
9.
Place the bracket on the Ethernet
connector and mount the connector to the
rear panel.
10.
Mount the panel again on the rear of the
table console.
Note
. Strain relief should be added to the gantry-PC interface cable as follows:
Procedure Details
11.
After making all connections, take all
excess cable length to the interior of the
gantry.
12.
To prevent cable connections from being
stressed, attach two large Ty-Wraps to the
cable on the inside of the cabinet.
To mount the laptop power supply in the PC cart stand, proceed as follows:
Procedure Details
1.
Remove the cover on the rear side of the
stand.
Bracket