Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-25
Procedure Details
The table base is shipped in a second wooden
crate (31.5
×
47.2
×
36.6 inch / 80
×
120
×
93
cm).
3.
Remove the plywood sides and top.
4.
Remove the protection bag by carefully
cutting around the edges.
The acquisition laptop PC, system accessories,
covers and installation Dolly are all shipped
in a third wooden crate (31.5
×
57
×
34 inch /
80
×
145
×
87 cm).
5.
Remove the plywood sides and top.
6.
Remove the protection bag by carefully
cutting around the edges.
The patient table option is shipped in two
cardboard boxes.
The pallet and mattress are shipped in one
box:
•
19.7
×
110.6
×
5.9 inch / 50
×
281
×
15 cm
for the long table,
•
19.7
×
81.5
×
5.9 inch / 50
×
207
×
15 cm
for the short table.
The table end support is shipped in a second
box (31.5
×
23.6
×
31.5 inch / 80
×
60
×
80
cm).
7.
Open the boxes and remove the protection
bags by carefully cutting around the edges.
The laptop PC mounting option is shipped in a
single box.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......