5.
Unscrew the motor cable connections
from the motion controller connector:
Green to pin 4 (PE)
Red to pin 5 (M1)
White to pin 6 (M2)
Black to pin 7 (M3).
6.
Disconnect the spring from the motor and
Acme screw bracket.
7.
Use a 2.5 mm Allen key to unscrew the
rail on the underside of the line source
compartment.
This will loosen the bracket holding the motor
and ball screw.
8.
Carefully lift up the bracket and side panel
assembly.