Planned Maintenance
12-116
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
12.8
Planned Maintenance
12.8.1
Lubrication
12.8.1.1
Tools required
•
Grease: Klüber Microlube
®
GL261.
12.8.1.2
Acme Screw
The acme screw in the line source compartment must be greased every 6 months.
Procedure Details
1.
Use a brush to apply Microlube grease
along the entire length of the Acme screw
inside the line source compartment.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......