Software Update Procedures
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
7-21
Procedure Details
3.
Start the Windows Explorer and select the
CD-ROM drive.
4.
Select the directory
Firewire driver
version 1.5 for the stationary
PC
.
5.
Run the file
ubCore.exe
by double-
clicking on the name of the file.
6.
Follow the instructions given.
7.
After installing the driver, run the Fire-
Viewer software to check that FireWire
communication is working properly. In-
structions are provided in Chapter 5
Diag-
nostics
.
Note
. If you display the Device Manager (right-click the My Computer icon, select Properties,
click the Hardware tab and then the Device Manager button), the icon
Unibrain driver/
FireBoard400
(the name of the FireWire driver) is marked with a yellow Exclamation mark. In
principle, this means that the device is not working properly. Reinstalling the FireWire driver does
not solve the problem.
The reason for this is that there is a minor mismatch between Microsoft Windows 2000 and the
FireWire driver used by the CardioMD acquisition software. However, this does not affect Fire-
Wire functionality. To ascertain that the FireWire communication between the gantry and the ac-
quisition PC is working properly, run the FireViewer program that is installed along with the
driver. See instructions in Chapter 5
Diagnostics
.
Installing the FireWire driver on Windows XP
Note
. This section applies to acquisition PC’s running Windows XP.
Procedure Details
1.
Insert the CardioMD software CD in the
acquisition PC’s CD-ROM drive.
2.
Close all programs on the acquisition PC.
3.
Start the Windows Explorer and select the
CD-ROM drive.
4.
Select the directory
FireWire OHCI
driver version 3.0 for laptop
PC
.
5.
Run the file
ubCore3_23.exe
by double-
clicking on the name of the file.
6.
Follow the instructions given.
7.
After installing the driver, run the Fire-
Viewer software to check that FireWire
III
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......