Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-13
8.2
Shipping, Handling, Installation and Storage
8.2.1
Shipping Containers and Weights
Prior to installation, either the customer or a structural engineer must qualify that the proposed
room location is capable of supporting a CardioMD system. Furthermore, the entire equipment
path from the receiving dock to the clinical room must also be qualified. This includes all freight
elevators if applicable.
The system is shipped with either a short or a long patient table in 6 containers. Table 8.3 lists size
and weight.
Shipping box with
Shipping weight
Size W
×
D
×
H
Gantry incl. detector and collimators
1653 lbs.
750 kg
34
×
61
×
63 inch
86
×
155
×
160 cm
Table base
507 lbs.
230 kg
31,5
×
47.2
×
36,6 inch
80
×
120
×
93 cm
Accessories, laptop PC, covers, dolly etc.
397 lbs.
180 kg
31,5
×
57
×
34,3 inch
80
×
145
×
87 cm
Option: Short table and mattress
35 lbs.
16 kg
19,7
×
81.5
×
5.9 inch
50
×
207
×
15 cm
Table support for short table option
66 lbs.
30 kg
31.5
×
23.6
×
31.5 inch
80
×
60
×
80 cm
Option: Long table and mattress
44 lbs.
20 kg
19.7
×
110.6
×
5.9 inch
50
×
281
×
15 cm
Table support for long table option
66 lbs.
30 kg
31.5
×
23.6
×
31.5 inch
80
×
60
×
80 cm
Laptop cart-mount option
110 lbs.
50 kg
31.5
×
47
×
37 inch
80
×
120
×
93 cm
Laptop wall-mount option
110 lbs.
50 kg
19.7
×
110.6
×
10.6 inch
50
×
281
×
27 cm
Laptop desk-mount option
110 lbs.
50 kg
19.7
×
110.6
×
10.6 inch
50
×
281
×
27 cm
Collimator cabinet
110 lbs.
50 kg
18.9
×
43.3
×
22.4 inch
48
×
110
×
57 cm
AC Option
122 lbs.
55 kg
15.7
×
31.5
×
25.6 inch
40
×
80
×
65 cm
Table 8.3 Shipping containers, contents, weight and dimensions
8.2.2
Passageway Clearances
Check all doors, corridors and elevators to ensure adequate clearances for the moving of the
equipment.
The primary method for transporting the CardioMD gantry and table base from the receiving dock
to the clinical room is a special dolly supplied with the system. A standard pallet jack best
transports the table and remaining packages.
Below are listed the minimum passage clearances required for moving a CardioMD system.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......