Installation
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
8-19
8.2.6.2
Cabling
The length of the CardioMD power cord is 4 m (13 ft).
The acquisition PC cable comes from the right-hand side of the table console. The length of this
cable depends on the acquisition PC option: Cart-mount, wall-mount or desk-mount.
Cart-mount Wall-mount Desk-mount
Figure 8.12 Acquisition PC and gantry cable lengths
8.2.7
CardioMD Floor Levelness
The floor occupied by the CardioMD system must be in level within ½ inch (12.5 mm).