Repair Procedures
12-62
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
19.
Mount the two Allen screws holding the
side rail to the compartment.
20.
Mount the two Allen screws holding the
source motion board.
Note
. It is important to fasten the side rail
before mounting the Allen screws in the
source motion board.
21.
Mount the actuator bracket, inserting the
anchor into the solenoid.
Note
. Make sure that the end of the bracket
fits into the groove in the side of the line
source compartment.
22.
Re-insert the shutter indicator / manual
lever.
23.
Mount the four Allen screws holding the
indicator and the actuator bracket to the
slide bearing.
Note
. Do
not
tighten the Allen screws yet.
Summary of Contents for CARDIOMD