Detailed Description
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
3-25
DYNCOR
LOW LEVEL
OUTPUT SIGNAL
INPUT SIGNAL
DYNCOR HIGH LEVEL
The nonlinear transfer function is characterized
by two break points, HIDYN and LODYN.
The levels of these break points are always
proportional to the energy of the events. The
relative levels are set by a value in the tuning
tables. These values are factory set and not
subject to field adjustments