Detector Calibration
4-28
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Procedure Details
15.
Click Exit to close the Uniformity Calibra-
tion page.
16.
Replace the FOV mask with the collima-
tor.
17.
Slide the collimator on the other detector
back in place.
18.
Press the hand controller Collision Over-
ride button and wait for a minimum of 4
seconds.
19.
Repeat the procedure outlined above to
calibrate the other detector.
20.
Click Exit to close the Uniformity Calibra-
tion page.
Extrinsic Uniformity Calibration Procedure
Procedure Details
1.
Ensure that the CardioMD system is in the
patient load position and that the acquisi-
tion PC is displaying the Persistence page.
2.
Click the Tools button.
The Calibration & Configuration menu ap-
pears.
3.
In the Calibration & Configuration menu,
click Uniformity calibration.
The Uniformity Calibration page appears.
4.
Specify the energy settings for the uni-
formity correction table to be acquired.
5.
Select Extrinsic.
6.
Select the detector to be calibrated.
The uniformity correction table is assigned a
filename of the form:
U.coll.ew1.ew2.ew3.detectorNum
Where:
coll
is set to
LEGP
,
LEHR
or
NONE
, depending
on the collimator mounted on the detector.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......