Detailed Description
12-26
CardioMD Service Manual
1MAN0158-H11 04-2005 CSIP Level 1
Detector
Move line source to
Command sequence
-> canmsg 533 0x80 0
-> canmsg 533 0x06 0
-> canmsg 533 0x0F 0
-> canmsg 533 0x0F 0
-> canmsg 1301 0 0xc0 3 0 0 0xc0 3 0
400 mm
(position furthest
from motion
controller)
-> canmsg 1045 0xaa 0xaa 0x10 0 0xaa 0xaa 2 0
-> canmsg 533 0x1f 0
-> canmsg 533 0x6f 0
Table 12.5 Commands for manual positioning of line source housing
Note
. In Table 12.5, some of the commands to be sent to the line source motion controller are
printed on a gray background. These commands serve to reset the motion controller and need only
be sent to the motion controller once.
Solenoid Driver
The driver for the solenoid activating the shutter (see Figure 12.18, page 12-20) is a switch-mode
type, powered from the E-Stop controlled 48 V supply. For the first 0.5 second after the driver is
activated, it operates at 100 % duty cycle. After this, it operates at 27 % duty cycle. The solenoid
driver is activated from an output port on the line source motion controller.
12.3.5
CardioMD Detector and Table Console
When the CardioMD AC Option is installed, the CardioMD detector and table console are
provided with additional 3 dedicated circuit boards, a 48 V PSU and associated cables. See AC
installation chapter for details.
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......