CardioMD AC Option
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
12-57
Procedure Details
When all of the above steps have been
performed, remove the collision detection
shutter end cover:
5.
Use a T08 Torx key to remove the 3
screws indicated in the photo on the right.
6.
Remove the cover.
7.
Disconnect the solenoid connector from
the AC source motion board.