Calibration
1MAN0158-H11 04-2005 CSIP Level 1
CardioMD Service Manual
4-23
Procedure Details
3.
On the Persistence page, click the Tools
button.
4.
When the Calibration & Configuration
menu appears, click Service.
5.
If prompted, enter the service password
(see page 4-9 for instructions).
6.
In the Service Calibrations menu, click
LinCor.
The LinCor page appears.
7.
In the Counts field, enter 60000000 (60
million).
8.
Set Width to 30.
9.
Select the appropriate isotope.
10.
Select the detector to be calibrated.
11.
Click Continue.
12.
Press and hold the hand controller Colli-
sion Override button for a minimum of 4
seconds before proceeding as follows:
13.
While still pressing Collision Override,
click OK to accept automated motions
bringing the system into position for cali-
bration of the selected detector.
Note
. The hand controller Collision Override
button must be kept depressed during all
automated motions to enable power to the
motors.
Note
. Automated motions must be started
from the patient load position.
14.
Place the X-linearity phantom on the de-
tector.
Note
. The X-linearity phantom has slits along
its short axis.
X-linearity phantom
Summary of Contents for CARDIOMD
Page 1: ...CARDIOMD SERVICE MANUAL 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 10: ...Introduction x CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 16: ...Safe Operating Practices 1 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 164: ...Detector Calibration Files 5 60 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 328: ...Installation Procedures 9 38 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 348: ...Planned Maintenance Procedures 11 6 CardioMD Service Manual 1MAN0158 H11 04 2005 CSIP Level 1 ...
Page 465: ......